Instruction Manual PCD 650
7
1.5 Connecting Peripherals
1.5.1 Probes (pH/Conductivity/DO)
Attach the probes with correct type of connectors as indicated in the figure
below.
Figure 6 : Connecting ph, Cond and DO Probes
The PCD 650 meters allows to connect a pH or ISE,
DO and a conductivity probe at the same time. Two
types of conductivity electrodes: the 2-cell
conductivity cell and the 4-cell conductivity cell can
be used with this meter and only one can be
connected at a time. The meter can also adjust cell
constant from 0.010 to 10.000. This electrode
features a built-in temperature sensor for Automatic
temperature compensation (ATC). It has a specially
designed housing that provides fast temperature response
and reduces air bubble entrapment, which makes it easy to
obtain accurate, stable readings.
Proper use of electrode is essential to ensure that the optimum measurement
is taken in a short time. The removable protective plastic electrode guard on
the conductivity cell is meant for simple periodic maintenance and it must be
kept in tact during measurement and calibration. Always immerse the
electrode beyond upper steel band as shown in Figure 7. We recommend
that you do not submerge the electrode above the protective yellow cap. You
can submerge the cable for brief periods of time, but not continuously.
This meter uses a special notched 6-pin connector to attach the DO probe to
the meter. The Dissolved Oxygen Probe works on galvanic principle, that is, it
pH Probe (BNC)
Conductivity Probe
with built in
temperature sensor
(8-pin Connector)
Figure 7: Probe
immersion level
DO Probe with
built in temperature
sensor (6-pin
Connector)