Alpha pH/ORP 1000 Process Controller/Transmitter

Página 2

5 3.2 Back Panel ENSURE electrical mains is disconnected before proceeding. The back panel consists of two connectors. The first connector is the 17

Página 3 - Preface

64 OVERVIEW 4.1 Keypad and Display 4.1.1 Keypad • Perform rapid 2-point calibration • Allows entry to Set up mode • Select individual functions

Página 4 - Safety Information

7 SETUP HOLD 000S.Cd4.2 Function Groups The main function and sub-function groups are organised in a matrix format for configuration and selection of

Página 5 - TABLE OF CONTENTS

84.3 Control Concept The main function and sub-function groups are organised in a matrix format as shown below. These functions can be accessed via

Página 6 - 1 INTRODUCTION

9 5 MEASUREMENT 5.1 Display in Measurement mode When the controller is initially powered on, it automatically enters into the Measurement mode afte

Página 7 - 2.2 Unit Dimensions

105.2.1 How to enter and change parameters in Calibration mode 1) Press the CAL key. The upper display shows “000” and the lower display shows “C.C

Página 8 - 3 ELECTRICAL CONNECTIONS

11 HOLD 6.96 7.00 4.024.00 SETUP HOLD CAL 59.0 mV 7.00 pH READY HOLD CAL CALpHSETUP HOLD pH pH pH pH 6 CALIBRATION You can reach the Calib

Página 9 - Matching Pin (PMP)

12CALSETUPHOLD226SETUPHOLDC8) Press the ENTER key. If you entered the calibration mode using the CAL key, the controller will return to the Measurem

Página 10 - FUSE 250VAC

13 6.3 ORP – % Calibration The ORP % mode allows you to calibrate at two points: a low concentration sample (20%) and a high concentration sample (80

Página 11 - 4 OVERVIEW

147 ADVANCED SET UP MODE 7.1 Electrode Offset (OFS) sub-function You can perform electrode offset only in the pH mode. This mode allows you to chan

Página 12 - 4.2 Function Groups

R AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink,

Página 13 - 4.3 Control Concept

15 7.2.2 Temperature calibration (ATC mode only) 1) Select “ATC on” as described above in Section 7.2.1. 2) Press ENTER key. The upper display ind

Página 14 - 5 MEASUREMENT

167.3 Control Relay A/ Control Relay B (SP1/SP2) sub-function The SP1 option sets operating parameters for Relay A; the SP2 option sets operating pa

Página 15

17 1) Follow directions in 7.3.1 to enter Control Relay mode. 2) Press the ENTER key. Scroll with the ▲ or ▼ key until the upper display shows the h

Página 16

187.4 Controller (Cntr) sub-function You can set the controller’s parameters in this sub-function. 7.4.1 Entering Controller sub-function 1) Ent

Página 17 - adjustment

19 1) Follow directions in 7.4.1 to enter Controller mode. 2) Press the ENTER key. Scroll until the lower display shows “rEL” and the upper display

Página 18

207.5 Current Output (rng) sub-function This sub-function lets you set the controller current output range of this unit. 7.5.1 Entering current ou

Página 19 - 7 ADVANCED SET UP MODE

21 7.6 Configuration (ConF) sub-function This group of parameters lets you configure the controller to suit your requirements. 7.6.1 Entering the Co

Página 20

227.6.3 Selecting Alarm or Wash function This function allows you to use the Alarm relay as Wash contact. The Wash contact is used in combination wi

Página 21

23 7.6.6 Configuring the Wash contact if relay 3 has been set to the Wash mode 1) Follow direction in 7.6.1 to enter Configuration mode. 2) Press

Página 22

247.6.8 Reverting to factory default settings Use this parameter to reset all settings to factory default. 1) Follow directions in 7.6.1 to enter C

Página 23

Preface This manual serves to explain the use of the Alpha pH1000 pH/ORP process controller/transmitter. The manual functions in two ways, firstly a

Página 24

25 8 AUTO/MANUAL MODE Regardless of the mode, you can control devices connected to Relay A or Relay B from the front panel of this controller. In Auto

Página 25

269 TECHNICAL SPECIFICATIONS pH Range 0.00 to 14.00 pH Resolution 0.01 pH Relative Accuracy ± 0.01 pH mV Range 0 to 100.0% / -1000 to +

Página 26

27 Mechanical Specifications Dimensions (control panel housing - L x H x W) 175 x 96 x 96 mm Weights (control panel housing) max. 0.7 kg Material

Página 27

2810 ACCESSORIES Replacement Unit Product Description Thermo Scientific Code no. Eutech Code no. Alpha-pH1000 pH/ORP Controller/Transmitter with 0/

Página 28

29 11 GENERAL INFORMATION 11.1 Warranty Thermo Scientific warrants this product to be free from significant deviations in material and workmanship

Página 29

3012 APPENDICES 12.1 Appendix 1 Before opening the unit, ENSURE that the power cable is physically separated from the power supply. Jumper Positio

Página 30 - 8 AUTO/MANUAL MODE

31 12.2 Appendix 2 The following table shows the various pH values at different temperature of the solution during calibration. Temperature (oC) pH

Página 31 - 9 TECHNICAL SPECIFICATIONS

3212.3 Appendix 3 Simple Explanation on the Function of Hysteresis The controller relay activates when the set-point is reached. In the reverse dir

Página 32 - Mechanical Specifications

33 12.4 Appendix 4 General Instructions Concerning Controller Setting Xp0%50%100% Yh-Xw+XwXpProp. band Prop. bandSP 2SP 1MIN function MAX function C

Página 33 - 10 ACCESSORIES

34 The output relay of the pulse length controller is clock-timed. The switching period T remains constant. Depending on the divergence from the limi

Página 34 - 11 GENERAL INFORMATION

Safety Information The Controller/ Transmitter shall be installed and operated only in the manner specified in this Instruction manual. Only skilled

Página 37 - 4 . 0 4 . 5 9 . 5 1 0 . 0

Water Analysis InstrumentsNorth America166 Cummings CenterBeverly, MA 01915 USAToll Free: 1-800-225-1480Tel: 1-978-232-6000Dom. Fax: 1-978-232-6015Int

Página 38 - 12.4 Appendix 4

TABLE OF CONTENTS 1 INTRODUCTION ___________________________________________________ 1 1.1 Description of Unit...

Página 39

1 1 INTRODUCTION 1.1 Description of Unit Thank you for purchasing the ¼ DIN Alpha pH1000 pH/ORP process controller/transmitter. This unit is used for

Página 40

22 ASSEMBLY AND INSTALLATIONS 2.1 Measurement and Control System A typical measurement system consists of: • a pH/ORP process controller • a pH

Página 41

3 3 ELECTRICAL CONNECTIONS 3.1 Connection Diagram ENSURE electrical mains is disconnected before proceeding. Signal OutputHold InputAlarmRelay BRe

Página 42 - Water Analysis Instruments

4 Symmetrical high-impedance connections Asymmetrical connections Measuring CablepH CombinationElectrode