Instruction Manual CON 6/TDS 6
29
6.2 Select Cell Constant
This meter lets you select a cell constant of K = 1.0, 10,
or 0.1.
Use a cell of K = 1.0 for midrange
measurements
Use a cell of K = 10 for high range
measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range
measurements (below 20 µS or 10 ppm).
The cell included with your meter has a cell constant of
K = 1.0.
1) Enter the advanced setup as described in
Section 6.1.
2) Press or key until [CELC] appears on the
LCD. Press ENTER key.
3) Press or key to select either “1.0”, “0.1” or
“10.0”. Ensure the cell constant selected
correspond with the conductivity electrode you are
using.
4) Press ENTER key to select. The meter will take
you back to the menu, [CELC].
5) Press or key to move to the next menu or
press CAL to exit to measurement mode.