Press [ENTER] and the transmitter will ask to ‘Save Calibration’.
Select Yes to save the calibration values and the successful Calibration is confirmed on the
display.
The transmitter will return to the operation mode in approximately 5 seconds.
The above calibration process can then be repeated for Channel B of dual channel transmitters.
11.2.2 Calibrate Analog
Select the Analog Output you wish to calibrate. Each Analog output can be calibrated at 4 and
20 mA.
Connect an accurate milliamp meter to the Analog output terminals and then adjust the five digit
number in the display until the milliamp meter reads 4.00 mA and repeat for 20.00 mA.
As the five digit number is increased the output current increases and as the number is de-
creased the output current decreases. Thus coarse changes in the output current can be made
by changing the thousands or hundreds digits and fine changes can be made by changing
the tens or ones digits.
Pressing the [ENTER] key after entering both values will bring up a con-firmation screen.
Selecting No will discard the entered values, selecting Yes will make the entered values the
current ones.