Transmitter Cond/Res 2-Channel M300 34
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
Table 8.1: USP Section <645> Stage 1, EP WFI (Water for Injection), and EP Highly Purified
Water Conductivity Limits as a Function of Temperature
Temperature (ºC) Conductivity Limit (µS/cm)
0 0.6
5 0.8
10 0.9
15 1.0
20 1.1
25 1.3
30 1.4
35 1.5
40 1.7
45 1.8
50 1.9
55 2.1
60 2.2
65 2.4
70 2.5
75 2.7
80 2.7
85 2.7
90 2.7
95 2.9
100 3.1
Table 8.2: EP PW (Purified Water) Conductivity Limits as a Function of Temperature
Temperature (ºC) Conductivity Limit (µS/cm)
0 2.4
10 3.6
20 4.3
25 5.1
30 5.4
40 6.5
50 7.1
60 8.1
70 9.1
75 9.7
80 9.7
90 9.7
100 10.2