EUTECH INSTRUMENTS ALPHA DO 500 DISSOLVED OXYGEN Manual de Instruções Página 28

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 49
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 27
Instruction manual αlpha DO 2000
23
1. Select the “CNTR” subfunction, then press the ENT key.
2. Selecting the controller type: press the or key to select the suitable controller type:
OFF = controller off
Use control Off to operate controller as a monitor only or to prevent relays from
switching.
L.CT = limit value control (on/off control).
Use limit control with pumps or valves for fast response
P/PI = proportional/integral control
Use proportional control to operate your pumps smoothly or for precise control of
proportioning valves. Use PI controller to eliminate steady state error.
Note: please refer to Appendix 4 for detailed information on controller settings.
Press the ENT key to confirm your selection.
If the controller is set to limit value control (L.CT):
3. Selecting the relay status under Non-Alarm condition: press the or key to choose
the desired relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to
confirm your selection.
If the controller is set to proportional control (P/PI):
4. Selecting the proportional controller type: press the or key to select the suitable
controller type (“PLC” = pulse length control, “PFC” = pulse frequency control). Press the ENT
key to confirm your selection.
If the proportional controller type is set to pulse length control (PLC):
5. Selecting the relay status under Non-Alarm condition: press the or key to choose
the desired relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to
confirm your selection.
6. Setting the proportional range: Press the or key to set the proportional range (setting
range: 10 to 500%). Press the ENT key to confirm your setting.
7. Setting the pulse length: Press the or key to set the pulse length (setting range: 0.5 to
20 seconds). Press the ENT key to confirm your setting.
8. Setting the integral action time: Press the or key to set the integral action time
(setting range: 0.0 to 999.9 minutes). Press the ENT key to confirm your setting.
Vista de página 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 48 49

Comentários a estes Manuais

Sem comentários