Instruction manual αlpha DO 2000
21
5. Setting the on-delay time lag: press the or key to enter the on-delay time for set point
1 (set point 2). The controller will delay activation of the relay for the number of seconds (0 to
2000 seconds) you select. Press the ENT key to confirm your setting.
Note: You can set a time delay for each relay, which stops the relay from switching
on the moment the set point is exceeded. This controller lets you set a 0 to 2000
seconds time delay before your relay activates.
6. Setting the off-delay time lag: press the or key to enter the off-delay time for set point
1 (set point 2). Your controller will delay deactivation of the relay for the number of seconds (0
to 2000 seconds) you select. Press the ENT key to confirm your setting.
Note: You can set a time delay for each relay, which stops the relay from switching
off the moment the value reaches the set point and hysteresis. This controller lets
you set a 0 to 2000 seconds time delay before your relay deactivates.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).