EUTECH INSTRUMENTS ECOSCAN TEMP JKT PORTABLE METER Manual de Instruções Página 1

Consulte online ou descarregue Manual de Instruções para Medição, teste e controlo EUTECH INSTRUMENTS ECOSCAN TEMP JKT PORTABLE METER. EUTECH INSTRUMENTS ECOSCAN TEMP JKT PORTABLE METER Instruction manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
Technolo
g
y
M
ade
E
a
s
y
.
.
.
Instruction Manual
Temp JKT
Temperature Meter Series
68X243615
Rev. 2 11/03
Vista de página 0
1 2 3 4 5 6 ... 19 20

Resumo do Conteúdo

Página 1 - Temp JKT

TechnologyMadeEasy...Instruction Manual Temp JKT Temperature Meter Series 68X243615 Rev. 2 11/03

Página 2

76. Press ENTER key to confirm calibration. 7. The LCD displays “CO” momentarily, and the meter then reverts to measurement mode. See figure below.

Página 3 - TABLE OF CONTENTS

8 4. MEASUREMENT 4.1 Taking Measurements 1. Power on the meter. The meter automatically enters Temperature mode. The “J”, “K” or “T” annunciator wil

Página 4 - 1. INTRODUCTION

94.5 Maximum and Minimum Hold Mode With the Maximum and Minimum Hold mode, the Temp JKT meter can be used as a maximum registering (or minimum regis

Página 5 - 2. GETTING STARTED

105. DISABLE AUTO POWER OFF FEATURE By default this meter will auto power-off 17 minutes after last key operation. This is to conserve battery power

Página 6

11 6. PROBE CARE AND MAINTENANCE For best results, always clean the temperature probe with clean tissue paper after measurement to keep the probe i

Página 7 - 2.4 Inserting New Batteries

128. SPECIFICATIONS Temperature Range Type J Type K Type T -200 to 1000 °C (-328 to 1832 °F) -250 to 1372 °C (-418 to 2502 °F) -250 to 400 °C (-418

Página 8 - 2.7 Switching the Meter On

139. ACCESSORIES Replacement Meter and Meter accessories Item Eutech Instruments Ordering Code Oakton Instruments Ordering Code Temp JKT meter onl

Página 9 - 3. CALIBRATION

14 10. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from

Página 10 - From measurement mode

15 11. RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any

Página 12

Preface This manual serves to explain the use of the Temp JKT temperature meter. This manual functions in two ways: first, as a step by step guide t

Página 13

For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our webs

Página 14 - 7. TROUBLESHOOTING

TABLE OF CONTENTS 1. INTRODUCTION 1 1.1 Introducing the Temp JKT 1 2. GETTING STARTED 2 2.1 Description of Keypad Functions 2 2.2 Description

Página 15 - 8. SPECIFICATIONS

11. INTRODUCTION 1.1 Introducing the Temp JKT Thank you for purchasing the Temp JKT meter. This economical microprocessor-based handheld meter works

Página 16 - 9. ACCESSORIES

22. GETTING STARTED 2.1 Description of Keypad Functions Temp JKT meter has seven keys on its splash-proof keypad. These keys include ON / OFF, CAL,

Página 17 - 10. WARRANTY

3 2.3 Inserting & Removing the Rubber Boot 1. To remove meter from rubber boot, push out from the bottom edges of meter until it is completely

Página 18 - 11. RETURN OF ITEMS

4 2.4 Inserting New Batteries The battery compartment is found at the back of instrument. To open the battery compartment, push in the direction of

Página 19 - NOTES

52.5 Battery Replacement A “LO” annunciator in the LCD alerts you when battery power is running low. Replace with the same type as recommended by the

Página 20

63. CALIBRATION 3.1 Temperature Calibration The temperature sensor included with your meter is factory calibrated. Over time, the temperature calibra