EUTECH INSTRUMENTS ALPHA PH1000 PHORP CONTROLLERTRANSMITTER Manual de Instruções Página 30

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 29
Instruction Manual alpha-pH 1000
30
9 TECHNICAL SPECIFICATIONS
pH Range 0.00 to 14.00 pH
Resolution 0.01 pH
Relative Accuracy +/- 0.01 pH
mV Range 0 to 100.0 % / - 1000 to +1000 mV
Resolution 0.1 % / 1 mV
Relative Accuracy +/- 1 mV
Temperature - 9.9 to + 125.0
o
C
Resolution 0.1
o
C
Relative Accuracy +/- 0.5
o
C
Sensor Pt 100 /Pt 1000 (jumper selectable)
Temperature Compensation Auto / manual (reference at 25.0
o
C)
Set-point and Controller Functions
Function (switchable) limit controller
pulse length/frequency controller
Controller characteristics proportional controller
Adjustable period with pulse length controller 0.5 to 20 sec.
Adjustable period with pulse frequency controller 60 to 120
pulses/min
Pickup / Dropout delay 0 to 2000 sec.
Switching pH hysteresis 0.1 to 1 pH
Switching ORP hysteresis 1 to 10.0 % / 10 to 100 mV
Contact outputs, controller 2 potential-free change-over contacts
Switching voltage / current / power max. 250 VAC / max. 3A / max. 600 VA
Alarm Functions
Function (switchable) Latching / pulse
Wash cycle 0.1 to 199.9 hrs
Wash duration 1 to 1999 s
Pickup delay 0 to 2000 s
Switching voltage / current / power max. 250 VAC / max. 3A / max. 600 VA
Electrical Data and Connections
Power Requirements 110 / 220 VAC (jumper selectable)
Frequency 48 to 62 Hz
Power requirements Max. 7 VA
Signal Output 0/4 to 20 mA, galvanically isolated
Load
max. 600
pH / ORP input BNC, screw terminals
Connection terminal Terminal blocks 5-pole / 17-pole, removable
Mains fuse / fine wire fuse slow-blow 250 V / 100 mA
EMC Specifications
Emissions According to EN 50081-1
Susceptibility According to EN 50082-1
Environmental Conditions (For Indoor Use)
Altitude Up to 2000 m
Maximum Relative Humidity 80% up to 31
o
C decreasing linearly to 50% at 40
o
C
Power Supply Mains supply voltage fluctuations not exceeding +
10% of the nominal voltage
Transient Overvoltages Category II
Pollution Degree Category II
Ambient temp. operating range -10 to +50
o
C
Vista de página 29