EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER (REVISION 3) Manual de Instruções Página 28

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 27
28
8.7 Viewing TDS Calibration Data – P5.0
Program 5 is a “view only” option, which displays the value at which calibration was
performed.
P5.1: view Calibrated TDS value (if no calibration performed for this range, controller
displays ‘- - - - ‘.
These parameters will change each time you recalibrate the controller.
P5.0
C L
P5.1
CAL
300
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and return
to Measurement mode.
P5.0: Viewing Calibrated TDS data
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P5.0 and CAL. Press ENTER.
3. Screen will scroll P5.1, ‘CAL’, then Calibrated value (in ppm or ppt).
4. Press and keys together (ESCAPE) twice, to return to measurement mode.
Vista de página 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 39 40

Comentários a estes Manuais

Sem comentários