Instruction Manual αlpha-pH2000D
- 18 -
MEAS
pH
O
C
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
SE TU P
HOLD
SETUP
HOL D
SETUP
HO LD
SETUP
HOL D
SETUP
HO LD
SETUP
HOL D
SE TUP
HOLD
SETUP
HO LD
SETUP
HO LD
SETUP
HO LD
Temperature settings
see section 6.3
Relay A (set point 1) settings
see section 6.4
Relay B (set point 2) settings
see section 6.4
Controller settings
see section 6.5
Current output 1 settings
see section 6.6
Current output 2 settings
see section 6.7
Wash contact settings
see section 6.8
Unit settings
see section 6.9
Calibration
see section 5
6 SETUP MODE
6.1 Enter Setup mode
In the Setup mode the transmitter can be configured to your individual requirements.
1. While in Measurement mode press the ENT key.
2. The display prompts you to enter the security code. Set the security code with
or key to:
“SCD 22” if you want to change parameter settings
“SCD 000” if you want to view only parameter settings (view only mode)
3. Press the ENT key.
Note: Note: to exit Setup mode at any time press the and key simultaneously
(escape). The transmitter returns to the Measurement mode.