11. SPECIFICATIONS
DO % Saturation
Range
Resolution
Relative accuracy
Offset adjustment
Calibration Points
0 to 300%
0.1 % (0.0 to 199.9%); 1% (200 to 300%)
± 0.5% of value
± 10.0% of reading
(2 points) 100% in saturated air or air-saturated water,
0% in zero oxygen solution
DO mg/L or ppm
Range
Resolution
Relative accuracy
Calibration
0.0 to 30.0
0.01
± 0.5% of value
(1 point) ± 40% from the factory default value.
Minimum reading allowed is 2.00.
Temperature
Range
Resolution
Relative accuracy
Calibration
0.0 to 50.0 ºC / 32.0 to 122.0 ºF
0.1 ºC / ºF
± 0.5ºC / ± 0.9ºF
Offset in 0.1 º increments; Offset range: ± 5 ºC / 9 ºF
Salinity Correction
Range
Resolution
Method
0.0 to 50.0 ppt
0.1 ppt
Automatic correction after manual input
Pressure Correction
Range
Resolution
Method
450 to 825 mmHg / 0.6 to 1.1 bar
1 mm Hg / 0.1 bar
Automatic correction after manual input
Probe (DO / Temp) Galvanic / Thermistor
Avg. Response Time 60 seconds to achieve 95% of the reading
Hold Function Yes
Memory 100 data sets
Input BNC for DO & 2.5 mm phono plug for temperature
Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)
Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)