page | 90
PCD 650 PC 650 PD 650 CD 650 PD 300 PC 300 PC 10
Multi-Parameter
Handheld Meters
Speci cations
Measuring Parameter
pH / ORP / Ion /
Conductivity / TDS /
Salinity / Resistivity /
Dissolved Oxygen /
ºC / ºF
pH / ORP / Ion /
Conductivity / TDS /
Salinity / Resistivity /
ºC / ºF
pH / ORP / Ion /
Dissolved Oxygen /
ºC / ºF
Conductivity / TDS /
Salinity / Resistivity /
Dissolved Oxygen /
ºC / ºF
pH / Dissolved
Oxygen / ºC / ºF
pH / Conductivity /
TDS / ºC / ºF
pH /
Conductivity / ºC
Highlights Waterproof, GLP, RS232C, IrDA, multi-parameter display
W
at
erproof with
back lighting
Waterproof,
speciality probe
compatible
Waterproof,
combined multi-
submersible probe
pH
R
ange -2.000 t
o 20.000 pH – -2.00 to 16.00 pH 0.00 to 14.00 pH
Resolution 0.1 / 0.01 / 0.001 pH – 0.01 pH
Accuracy ±0.002 pH – ±0.01 pH
Cal. Points 1 (offset) to 6-points – 5 3
Auto-Buffer
Rec
o
gnition
Yes–Yes
Slope &
Off
set Display
Yes – Yes –
Ion
Range 0.001 to 19900 –
Resolution 2 or 3 digits –
Accuracy 0.5 % full scale (monovalent) ; 1 % full scale (divalent) –
Cal. Points Up to 8 –
ORP
R
ange ±2000.0 mV –
Resolution 0.1 mV –
A
ccuracy ±0.2 mV –
Temperature
Range
-10.0 to 110.0 ºC /
14.0 to 230.0 ºF
(met
er)
0 to 60.0 ºC /
32 to 140 ºF (for DO)
-10.0 to 110.0 ºC /
14.0 to 230.0 ºF
(meter)
-10.0 to 110.0 ºC / 14.0 to 230.0 ºF (meter)
0 to 60.0 ºC / 32 to 140 ºF (for DO)
-10.0 to 110.0 ºC / 14.0 to 230.0 ºF (meter)
0.0 to 80.0 ºC
(with supplied
probe)
Resolution 0.1 ºC / 0.1 ºF 0.1 ºC
Accuracy ±0.5 ºC / ±0.9 ºF 0.3 ºC ±0.5 ºC / ±0.9 ºF ±0.5 ºC
Conductivity
R
ange
.. to 2.000 µS/cm
2.000 to 300.0 µS/cm
300.0 µS t
o 4.000 mS
4.000 to 40.00 mS/cm
40.00 to 500.0 mS/cm
–
.. to 2.000 µS/cm
2.000 to 300.0 µS/cm
300.0 µS to 4.000 mS
4.000 to 40.00 mS/cm
40.00 to 500.0 mS/cm
–
.. to 19.99 / 199.9 /
1999 µS/cm ;
.. to 19.99 /
199.9 mS/cm
.. to 19.99 / 199.9 /
1999 µS/cm ;
.. to 19.99 mS/cm
Resolution
0.01µS/cm /
0.1 µS/cm /
0.001mS/cm /
0.01mS/cm /
0.1 mS/cm
–
0.01µS/cm /
0.1µS/cm /
0.001mS/cm /
0.01mS/cm /
0.1 mS/cm
–
0.01 / 0.1 / 1 µS/cm ;
0.01 / 0.1 mS/cm
0.01 / 0.1 / 1 µS/cm ;
0.01 mS/cm
Accuracy ±1 % full scale + 1 LSD – ±1 % full scale + 1 LSD
– ±
1 % full scale + 1 LSD
Cal. Points
4 (1 per range) auto,
5 (1 per range) manual
–
4 (1 per range) aut
o
,
5 (1 per range) manual
–
5 (1 per range)
manual
4 (1 per range)
manual
Cell Constant (K) 0.010 to 10.000 – 0.010 to 10.000 – 1.0
Temperature
Coefficient
Linear & pur
e – Linear & pure –
0.0 to 10.0 %
(adjustable)
2.00 %
Normalisation
Temperatur
e
15 to 30 ºC – 15 to 30 ºC –
15.0 to 30.0 ºC
(adjustable)
25 ºC
A
uto-Ranging Yes – Yes – Yes
TDS
Range
(Depending
On TDS
Factor)
.. to 2.000 ppm
2.000 ppm t
o 300.0 ppm
300.0 ppm t
o 4.000 ppt
4.000 ppt to 40.00 ppt
40.00 ppt to 500.0 ppt
–
.. to 2.000 ppm
2.000 ppm to 300.0 ppm
300.0 ppm to 4.000 ppt
4.000 ppt to 40.00 ppt
40.00 ppt to 500.0 ppt
–
.. to 9.99 / 99.9 / 999 ppm ;
.. to 9.99 / 99.9 ppt ;
Max of 199.9 ppt
based on
factor setting
–
Resolution
0.01 ppm /
0.1 ppm /
0.001 ppt /
0.01 ppt /
0.1 ppt
–
0.01 ppm /
0.1 ppm /
0.001 ppt /
0.01 ppt /
0.1 ppt
–
0.01 / 0.1 / 1 ppm ;
0.01 / 0.1 ppt
–
Accuracy ±1 % full scale + 1 LSD – ±1 % full scale + 1 LSD
–
±1 % full scale + 1 LSD –
Conversion
Fa
c
tor
0.40 to 1.0 – 0.40 to 1.0 – 0.40 to 1.0 –
Multi-Parameter
CyberScan
Waterproof Handheld
Meters Specifi cations
Comentários a estes Manuais