EUTECH INSTRUMENTS CYBERSCAN DO 5000 DO METER Manual de Instruções Página 22

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 76
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 21
Instruction Manual PD 300
5 MEASUREMENT
5.1 Taking pH Measurements
5.1.1
Automatic Temperature Compensation
Automatic Temperature Compensation only occurs when a compatible temperature sensing
element is plugged into the meter.
If there is no temperature sensor plugged into the meter, the default manual temperature
setting is 25 °C. You can manually set the temperature to match your working conditions
using a separate thermometer.
For automatic temperature compensation (ATC) simply
plug the temperature probe into the meter (see page 6
for directions). The ATC indicator will light up on the
LCD. See figure on right.
Figure 13: Automatic Temperature
Compensation
NOTE
The temperature sensing element must be in the
sample that you are calibrating or measuring.
5.1.2
17
Vista de página 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 75 76

Comentários a estes Manuais

Sem comentários